        | Title | : | Application of Photoacoustic Imaging for Pneumonia Detection | 
| Author | : | Caesarany Maqfiroh (1) Dr. drg. Rini Widyaningrum, M.Biotech (2) AHMAD MUJTAHID ANAS (3) Prof. Dr. Mitrayana, S.Si., M.Si. (4) | 
| Date | : | 0 2023 | 
| Keyword | : | acoustic, imaging, intensity,photoacoustic,pneumonia acoustic, imaging, intensity,photoacoustic,pneumonia | 
| Abstract | : | We used photoacoustic imaging (PAI) to visualize and compare acoustic intensity levels in pneumonia-affected and healthy chicken lungs. After histological confirmation of pneumonia, the samples were scanned and subjected to a 532- nm diode laser in a photoacoustic imaging system. The acoustic intensity level of pneumonia-affected tissue was examined and compared with that of healthy lung samples. The optimum laser frequency and duty cycle for imaging the samples were 17 kHz and 30%, respectively. The acoustic intensity levels of pneumonia-affected tissue and healthy lungs were −82.5 ± 1.8 dB and −79.9 ± 1.3 dB, respectively. We found that a simple PAI device consisting of a diode laser and condenser microphone could distinguish between pneumonia-affected and healthy lungs. Pneumonia-affected lungs produced less acoustic intensity than that healthy lungs, as supported by histological studies | 
| Group of Knowledge | : | Fisika | 
| Original Language | : | English | 
| Level | : | Internasional | 
| Status | : | Published | 
| No | Title | Action | 
|---|---|---|
| 1 | fullpaper.pdf 
                                                Document Type : [PAK] Full Dokumen 
                                             | View |