ACADSTAFF UGM

CREATION
Title : Development of low cost and accurate homemade sensor system based on Surface Plasmon Resonance (SPR)
Author :

FEBRILIAN DWI L (1) SUPARDIANNINGSIH (2) Muhammad Arifin, S.Si., M.Sc., Ph.D. (3) Prof. Dr. Kamsul Abraha (4)

Date : 1 2018
Keyword : - -
Abstract : In this paper, we developed homemade and computerized sensor system based on Surface Plasmon Resonance (SPR). The developed systems consist of mechanical system instrument, laser power sensor, and user interface. The mechanical system development that uses anti-backlash gear design was successfully able to enhance the angular resolution angle of incidence laser up to 0.01°. In this system, the laser detector acquisition system and stepper motor controller utilizing Arduino Uno which is easy to program, flexible, and low cost, was used. Furthermore, we employed LabView’s user interface as the virtual instrument for facilitating the sample measurement and for transforming the data recording directly into the digital form. The test results using gold-deposited half-cylinder prism showed the Total Internal Reflection (TIR) angle of 41,34°± 0,01° and SPR angle of 44,20°± 0,01°, respectively. The result demonstrated that the developed system managed to reduce the measurement duration and data recording errors caused by human error. Also, the test results also concluded that the system’s measurement is repeatable and accurate
Group of Knowledge : Fisika
Original Language : English
Level : Internasional
Status :
Published
Document
No Title Document Type Action
1 2018 Laksono_2018_J__Phys_%3A_Conf__Ser__1011_012043.pdf
Document Type :
View