Title : Load displacement simulation of CP-Ti/UHMWPE hip implant
Author :

Dr. Handoko, S.T., M.T. (1) Dr. Suyitno, S.T., M.Sc. (2) Ir. Rini Dharmastiti, M.Sc., Ph.D., IPM., ASEAN Eng. (3) dr. Rahadyan Magetsari, Sp.OT(K)., FICS., Ph.D. (4)

Date : 11 2021
Keyword : compressive load, simulation, finite element compressive load, simulation, finite element
Abstract : Hip implant made from titanium is special due to its lower modulus of elasticity to avoid stress shielding with the human bone. One type of load the material designed to withstand is a gradually increased compressive force which happened when the patient change his/her position from sitting to standing. This study examined the capability of a metal on polymer (MOP) implant made from commercially pure titanium (CP-Ti) and ultra-high molecular weight polyethylene (UHMWPE) loaded up to 3 kN according to ISO 14242 standard. Two CP-Ti/UHMWPE MOP models with femoral diameter of 22 mm and 32 mm were simulated with finite element. The results expressed in load displacement curves were validated with compressive load experimental tests. Both materials are capable to withstand the load. Simulation data are in good agreement with the experiments.
Group of Knowledge : Teknik Mesin (dan Ilmu Permesinan Lain)
Original Language : English
Level : Nasional
Status :
Published
Document
No Title Document Type Action
1 Bukti submit Handoko-Universitas Gadjah Mada Mail - [JMPB] Submission Acknowledgement.pdf
Document Type : Bukti Submitted
Bukti Submitted View
2 Handoko JMPB 2021 terbit 366-1434-3-PB.pdf
Document Type : Bukti Published
Bukti Published View
3 Handoko 2021 JMPB 366-1434-3-PB.pdf
Document Type : [PAK] Full Dokumen
[PAK] Full Dokumen View
4 Turnitin Load displacement simulation of CP-Ti_UHMWPE hip implant (1).pdf
Document Type : [PAK] Cek Similarity
[PAK] Cek Similarity View
5 Penelitian 2021 Handoko Load displcament JMPB Bukti korespodensi.pdf
Document Type : [PAK] Bukti Korespondensi Penulis
[PAK] Bukti Korespondensi Penulis View