        | Title | : | Pengembangan Sistem Deteksi Gas CO pada Gas Hembus Manusia Berbasis Spektroskopi ICOS | 
| Author | : | Restu Widiatmono (1) Julien Mandon (2) Frans J.M. Harren (3) Prof. Dr. Kusminarto (4) Dr. Moh. Ali Joko Wasono, M.S. (5) Prof. Dr. Mitrayana, S.Si., M.Si. (6) | 
| Date | : | 0 2014 | 
| Keyword | : | QCL, ICOS, absorption Spectroscopy, CO, exhaled breath detection QCL, ICOS, absorption Spectroscopy, CO, exhaled breath detection | 
| Abstract | : | An ultra sensitive exhaled breath gas detection system has been developed. DFB cw quantum cascade laser (QCL) centered at 4610 nm as a radiation source was used in combination with an off-axis integrated cavity output spectroscopy for carbon monoxide (CO) gas detection in a low gas pressure. a 15 cm long high-finesse cavity with an effective optical path length of 400 m has also been used as a sample gas cell. We demonstrated detection limit of 1 ppb with less than 2 second averaging time. 0,7 ppmv concentration of CO has been identified on a robust measurement of an exhaled breath detection. | 
| Group of Knowledge | : | Fisika | 
| Original Language | : | Bahasa Indonesia | 
| Level | : | Nasional | 
| Status | : | Published | 